جزئیات آسیبپذیری
CVE-2025-14096
Credential Disclosure vulnerability in Radiometer Products
شرح رسمی (انگلیسی)
A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system. Other related CVE's are CVE-2025-14095 & CVE-2025-14097. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required Configuration for Exposure: Attacker requires physical access to the analyzer. Temporary work Around: Only authorized people can physically access the analyzer. Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution. Exploit Status: Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.
امتیازهای CVSS
| امتیاز | نسخه | منبع | بردار |
|---|---|---|---|
| 8.4بالا | 3.1 | CVEProject cvelistV5 (CNA) (Radiometer) | CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H |
محصولات آسیبپذیر
| فروشنده | محصول | نسخهها | منبع |
|---|---|---|---|
| Radiometer Medical Aps | ABL800 BASIC and ABL800 FLEX Analyzers | Windows 7 Operating system | CVEProject cvelistV5 (CNA) |
| Radiometer Medical Aps | ABL800 BASIC and ABL800 FLEX Analyzers | Windows XP Operating system | CVEProject cvelistV5 (CNA) |
| Radiometer Medical Aps | ABL90 FLEX and ABL90 FLEX PLUS Analyzers | Windows 7 Operating system | CVEProject cvelistV5 (CNA) |
| Radiometer Medical Aps | ABL90 FLEX and ABL90 FLEX PLUS Analyzers | Windows XP Operating system | CVEProject cvelistV5 (CNA) |
| Radiometer Medical Aps | AQT90 FLEX Analyzers | Windows 7 Operating system | CVEProject cvelistV5 (CNA) |
| Radiometer Medical Aps | AQT90 FLEX Analyzers | Windows XP Operating system | CVEProject cvelistV5 (CNA) |
منابع و راهنماها
۱اطلاعات
- تاریخ انتشار
- ۱۴۰۴/۰۹/۲۶
- آخرین بهروزرسانی
- ۱۴۰۴/۰۹/۲۶
- وضعیت
- منتشرشده
- صادرکننده (CNA)
- Radiometer
پیشبینی بهرهبرداری (EPSS)
ارزیابی SSVC (CISA)
- قابل خودکارسازی
- خیر
- وضعیت بهرهبرداری
- ندارد
- اثر فنی
- کامل
ضعفها (CWE)
- CWE-798Use of Hard-coded Credentials
- CWE-250Execution with Unnecessary Privileges