Contec Health CMS8000 Patient Monitor
Exposure of Private Personal Information to an Unauthorized Actor vulnerability in Contec Health CMS8000 Patient Monitor
CVSS 4.0
8.2بالا
شرح رسمی (انگلیسی)
In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text patient data to a hard-coded public IP address when a patient is hooked up to the monitor. This could lead to a leakage of confidential patient data to any device with that IP address or an attacker in a machine-in-the-middle scenario.
امتیازهای CVSS
| امتیاز | نسخه | منبع | بردار |
|---|---|---|---|
| 8.2بالا | 4.0 | CVEProject cvelistV5 (CNA) (icscert) | CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:N/VA:N/SC:N/SI:N/SA:N |
| 5.9متوسط | 3.1 | CVEProject cvelistV5 (CNA) (icscert) | CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:N |
محصولات آسیبپذیر
| فروشنده | محصول | نسخهها | منبع |
|---|---|---|---|
| Contec Health | CMS8000 Patient Monitor | All versions | CVEProject cvelistV5 (CNA) |
منابع و راهنماها
۴- https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01
- https://www.cisa.gov/resources-tools/resources/contec-cms8000-contains-backdoor
- https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication
- https://www.bleepingcomputer.com/news/security/backdoor-found-in-two-healthcare-patient-monitors-linked-to-ip-in-china/
هشدار CVEهای بعدی این محصولمحصول را دنبال کنید تا آسیبپذیریهای تازه را در تلگرام بگیرید.
اطلاعات
- تاریخ انتشار
- ۱۴۰۳/۱۱/۱۱
- آخرین بهروزرسانی
- ۱۴۰۳/۱۱/۲۵
- وضعیت
- منتشرشده
- صادرکننده (CNA)
- icscert
پیشبینی بهرهبرداری (EPSS)
۰٫۷۷٪
احتمال بهرهبرداری در ۳۰ روز
۵۵٫۹٪
صدک (نسبت به همهی CVEها)
ارزیابی SSVC (CISA)
- قابل خودکارسازی
- خیر
- وضعیت بهرهبرداری
- ندارد
- اثر فنی
- جزئی
ضعفها (CWE)
- CWE-359Exposure of Private Personal Information to an Unauthorized Actor