CVE-2025-0626

Contec Health CMS8000 Patient Monitor

Hidden Functionality vulnerability in Contec Health CMS8000 Patient Monitor

CVSS 4.0
7.7بالا

شرح رسمی (انگلیسی)

The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device.

امتیازهای CVSS

امتیازنسخهمنبعبردار
7.7بالا 4.0 CVEProject cvelistV5 (CNA) (icscert) CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:P/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N
7.5بالا 3.1 CVEProject cvelistV5 (CNA) (icscert) CVSS:3.1/AV:N/AC:H/PR:N/UI:R/S:U/C:H/I:H/A:H

محصولات آسیب‌پذیر

فروشندهمحصولنسخه‌هامنبع
Contec Health CMS8000 Patient Monitor All versions CVEProject cvelistV5 (CNA)

منابع و راهنماها

۴
هشدار CVEهای بعدی این محصولمحصول را دنبال کنید تا آسیب‌پذیری‌های تازه را در تلگرام بگیرید.

اطلاعات

تاریخ انتشار
۱۴۰۳/۱۱/۱۱
آخرین به‌روزرسانی
۱۴۰۳/۱۲/۱۱
وضعیت
منتشرشده
صادرکننده (CNA)
icscert

پیش‌بینی بهره‌برداری (EPSS)

۱٫۱۴٪
احتمال بهره‌برداری در ۳۰ روز
۶۶٫۷٪
صدک (نسبت به همه‌ی CVEها)

ارزیابی SSVC (CISA)

قابل خودکارسازی
خیر
وضعیت بهره‌برداری
ندارد
اثر فنی
کامل

ضعف‌ها (CWE)

  • CWE-912
    Hidden Functionality