جزئیات آسیب‌پذیری

CVE-2024-9832

Baxter Life2000 Ventilation SystemBaxter life2000_ventilator_firmware

No limit on failed login attempts with Clinician Password or Serial Number Clinician Password on Life2000 Ventilator

CVSS 3.1
9.3بحرانی

شرح رسمی (انگلیسی)

There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.

امتیازهای CVSS

امتیازنسخهمنبعبردار
9.3بحرانی 3.1 CVEProject cvelistV5 (CNA) (Baxter) CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H

محصولات آسیب‌پذیر

فروشندهمحصولنسخه‌هامنبع
Baxter Life2000 Ventilation System 06.08.00.00 and prior CVEProject cvelistV5 (CNA)
Baxter life2000_ventilator_firmware v ≤ 06.08.00.00 CISA ADP / Vulnrichment

منابع و راهنماها

۱
هشدار CVEهای بعدی این محصولمحصول را دنبال کنید تا آسیب‌پذیری‌های تازه را در تلگرام بگیرید.

اطلاعات

تاریخ انتشار
۱۴۰۳/۰۸/۲۵
آخرین به‌روزرسانی
۱۴۰۳/۰۸/۲۸
وضعیت
منتشرشده
صادرکننده (CNA)
Baxter

پیش‌بینی بهره‌برداری (EPSS)

۰٫۱۷٪
احتمال بهره‌برداری در ۳۰ روز
۷٫۱٪
صدک (نسبت به همه‌ی CVEها)

ارزیابی SSVC (CISA)

قابل خودکارسازی
خیر
وضعیت بهره‌برداری
ندارد
اثر فنی
کامل

ضعف‌ها (CWE)

  • CWE-307
    Improper Restriction of Excessive Authentication Attempts