CVE-2024-48974
Baxter Life2000 Ventilation SystemBaxter life2000_ventilator_firmware
Life2000 Ventilator does not perform proper file integrity checks when adopting firmware updates
CVSS 3.1
9.3بحرانی
شرح رسمی (انگلیسی)
The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device's configuration settings and/or compromise device functionality by pushing a compromised/illegitimate firmware file. This could disrupt the function of the device and/or cause unauthorized information disclosure.
امتیازهای CVSS
| امتیاز | نسخه | منبع | بردار |
|---|---|---|---|
| 9.3بحرانی | 3.1 | CVEProject cvelistV5 (CNA) (Baxter) | CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H |
محصولات آسیبپذیر
| فروشنده | محصول | نسخهها | منبع |
|---|---|---|---|
| Baxter | Life2000 Ventilation System | 06.08.00.00 and prior | CVEProject cvelistV5 (CNA) |
| Baxter | life2000_ventilator_firmware | v ≤ 06.08.00.00 | CISA ADP / Vulnrichment |
منابع و راهنماها
۱
هشدار CVEهای بعدی این محصولمحصول را دنبال کنید تا آسیبپذیریهای تازه را در تلگرام بگیرید.
اطلاعات
- تاریخ انتشار
- ۱۴۰۳/۰۸/۲۵
- آخرین بهروزرسانی
- ۱۴۰۳/۰۸/۲۸
- وضعیت
- منتشرشده
- صادرکننده (CNA)
- Baxter
پیشبینی بهرهبرداری (EPSS)
۰٫۱۵٪
احتمال بهرهبرداری در ۳۰ روز
۴٫۵٪
صدک (نسبت به همهی CVEها)
ارزیابی SSVC (CISA)
- قابل خودکارسازی
- خیر
- وضعیت بهرهبرداری
- ندارد
- اثر فنی
- کامل
ضعفها (CWE)
- CWE-494Download of Code Without Integrity Check