جزئیات آسیب‌پذیری

CVE-2024-48966

Baxter Life2000 Ventilation SystemBaxter life2000_ventilator_firmware

Life2000 service tools for test and calibration do not support user authentication

CVSS 3.1
10.0بحرانی

شرح رسمی (انگلیسی)

The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.

امتیازهای CVSS

امتیازنسخهمنبعبردار
10.0بحرانی 3.1 CVEProject cvelistV5 (CNA) (Baxter) CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H

محصولات آسیب‌پذیر

فروشندهمحصولنسخه‌هامنبع
Baxter Life2000 Ventilation System 06.08.00.00 and prior CVEProject cvelistV5 (CNA)
Baxter life2000_ventilator_firmware v < 6.08.00.00 CISA ADP / Vulnrichment

منابع و راهنماها

۱
هشدار CVEهای بعدی این محصولمحصول را دنبال کنید تا آسیب‌پذیری‌های تازه را در تلگرام بگیرید.

اطلاعات

تاریخ انتشار
۱۴۰۳/۰۸/۲۵
آخرین به‌روزرسانی
۱۴۰۳/۰۸/۲۵
وضعیت
منتشرشده
صادرکننده (CNA)
Baxter

پیش‌بینی بهره‌برداری (EPSS)

۰٫۶۹٪
احتمال بهره‌برداری در ۳۰ روز
۵۳٫۱٪
صدک (نسبت به همه‌ی CVEها)

ارزیابی SSVC (CISA)

قابل خودکارسازی
بله
وضعیت بهره‌برداری
ندارد
اثر فنی
کامل

ضعف‌ها (CWE)

  • CWE-306
    Missing Authentication for Critical Function