جزئیات آسیبپذیری
CVE-2024-48966
Baxter Life2000 Ventilation SystemBaxter life2000_ventilator_firmware
Life2000 service tools for test and calibration do not support user authentication
CVSS 3.1
10.0بحرانی
شرح رسمی (انگلیسی)
The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.
امتیازهای CVSS
| امتیاز | نسخه | منبع | بردار |
|---|---|---|---|
| 10.0بحرانی | 3.1 | CVEProject cvelistV5 (CNA) (Baxter) | CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H |
محصولات آسیبپذیر
| فروشنده | محصول | نسخهها | منبع |
|---|---|---|---|
| Baxter | Life2000 Ventilation System | 06.08.00.00 and prior | CVEProject cvelistV5 (CNA) |
| Baxter | life2000_ventilator_firmware | v < 6.08.00.00 | CISA ADP / Vulnrichment |
منابع و راهنماها
۱
هشدار CVEهای بعدی این محصولمحصول را دنبال کنید تا آسیبپذیریهای تازه را در تلگرام بگیرید.
اطلاعات
- تاریخ انتشار
- ۱۴۰۳/۰۸/۲۵
- آخرین بهروزرسانی
- ۱۴۰۳/۰۸/۲۵
- وضعیت
- منتشرشده
- صادرکننده (CNA)
- Baxter
پیشبینی بهرهبرداری (EPSS)
۰٫۶۹٪
احتمال بهرهبرداری در ۳۰ روز
۵۳٫۱٪
صدک (نسبت به همهی CVEها)
ارزیابی SSVC (CISA)
- قابل خودکارسازی
- بله
- وضعیت بهرهبرداری
- ندارد
- اثر فنی
- کامل
ضعفها (CWE)
- CWE-306Missing Authentication for Critical Function