Vulnerability details

CVE-2025-30046

CGM CLININET

Remote Code Execution in CGM CLININET

CVSS 4.0
9.0Critical

Official description

In the CGM CLININET application server, multiple processes responsible for HL7 communication with external systems run with root privileges while processing untrusted data received from those external systems. An attacker in adjacent network able to send data to these HL7 communication processes can achieve remote code execution with root privileges.  The vulnerability was confirmed to affect versions up to 2026.MS3 but later versions may also be vulnerable.

CVSS scores

ScoreVersionSourceVector
9.0Critical 4.0 CVEProject cvelistV5 (CNA) (CERT-PL) CVSS:4.0/AV:A/AC:L/AT:P/PR:N/UI:N/VC:H/VI:H/VA:H/SC:H/SI:H/SA:H

Affected products

VendorProductVersionsSource
CGM CGM CLININET v ≤ 2026.MS3 CVEProject cvelistV5 (CNA)

References and advisories

2
Kolun on Telegram @kolunsec Exploited CVEs, key news and a daily digest Alerts for this product's next CVEsFollow the product to get new vulnerabilities on Telegram.

Details

Published
2026-10-23
Last updated
2026-10-28
State
Published
Assigner (CNA)
CERT-PL
Kolun priority
18 · Low

Exploit prediction (EPSS)

0.26%
Probability of exploitation in 30 days
18.2%
Percentile (among all CVEs)

SSVC assessment (CISA)

Automatable
No
Exploitation
None
Technical impact
Total

Weaknesses (CWE)

  • CWE-250
    Execution with Unnecessary Privileges