Vulnerability details

CVE-2025-0626

Contec Health CMS8000 Patient Monitor

Hidden Functionality vulnerability in Contec Health CMS8000 Patient Monitor

CVSS 4.0
7.7High

Official description

The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device.

CVSS scores

ScoreVersionSourceVector
7.7High 4.0 CVEProject cvelistV5 (CNA) (icscert) CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:P/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N
7.5High 3.1 CVEProject cvelistV5 (CNA) (icscert) CVSS:3.1/AV:N/AC:H/PR:N/UI:R/S:U/C:H/I:H/A:H

Affected products

VendorProductVersionsSource
Contec Health CMS8000 Patient Monitor All versions CVEProject cvelistV5 (CNA)

References and advisories

4
Kolun on Telegram @kolunsec Exploited CVEs, key news and a daily digest Alerts for this product's next CVEsFollow the product to get new vulnerabilities on Telegram.

Details

Published
2025-01-30
Last updated
2025-03-01
State
Published
Assigner (CNA)
icscert

Exploit prediction (EPSS)

1.14%
Probability of exploitation in 30 days
66.7%
Percentile (among all CVEs)

SSVC assessment (CISA)

Automatable
No
Exploitation
None
Technical impact
Total

Weaknesses (CWE)

  • CWE-912
    Hidden Functionality