Vulnerability details
CVE-2024-48966
Baxter Life2000 Ventilation SystemBaxter life2000_ventilator_firmware
Life2000 service tools for test and calibration do not support user authentication
CVSS 3.1
10.0Critical
Official description
The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.
CVSS scores
| Score | Version | Source | Vector |
|---|---|---|---|
| 10.0Critical | 3.1 | CVEProject cvelistV5 (CNA) (Baxter) | CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H |
Affected products
| Vendor | Product | Versions | Source |
|---|---|---|---|
| Baxter | Life2000 Ventilation System | 06.08.00.00 and prior | CVEProject cvelistV5 (CNA) |
| Baxter | life2000_ventilator_firmware | v < 6.08.00.00 | CISA ADP / Vulnrichment |
References and advisories
1
Kolun on Telegram @kolunsec
Exploited CVEs, key news and a daily digest
Alerts for this product's next CVEsFollow the product to get new vulnerabilities on Telegram.
Details
- Published
- 2024-11-15
- Last updated
- 2024-11-15
- State
- Published
- Assigner (CNA)
- Baxter
Exploit prediction (EPSS)
0.69%
Probability of exploitation in 30 days
53.1%
Percentile (among all CVEs)
SSVC assessment (CISA)
- Automatable
- Yes
- Exploitation
- None
- Technical impact
- Total
Weaknesses (CWE)
- CWE-306Missing Authentication for Critical Function