Vulnerability details

CVE-2024-48966

Baxter Life2000 Ventilation SystemBaxter life2000_ventilator_firmware

Life2000 service tools for test and calibration do not support user authentication

CVSS 3.1
10.0Critical

Official description

The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.

CVSS scores

ScoreVersionSourceVector
10.0Critical 3.1 CVEProject cvelistV5 (CNA) (Baxter) CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H

Affected products

VendorProductVersionsSource
Baxter Life2000 Ventilation System 06.08.00.00 and prior CVEProject cvelistV5 (CNA)
Baxter life2000_ventilator_firmware v < 6.08.00.00 CISA ADP / Vulnrichment

References and advisories

1
Kolun on Telegram @kolunsec Exploited CVEs, key news and a daily digest Alerts for this product's next CVEsFollow the product to get new vulnerabilities on Telegram.

Details

Published
2024-11-15
Last updated
2024-11-15
State
Published
Assigner (CNA)
Baxter

Exploit prediction (EPSS)

0.69%
Probability of exploitation in 30 days
53.1%
Percentile (among all CVEs)

SSVC assessment (CISA)

Automatable
Yes
Exploitation
None
Technical impact
Total

Weaknesses (CWE)

  • CWE-306
    Missing Authentication for Critical Function